BJ series of Disintegration testers are used for detecting disintegration of solid in prescriptive conditions.
Applicable standards:
National standard(Chinese codex)/2000/
Corporate standard(Disintegration tester)Q/12XQ0I88-2004
Feature:
There are eighteen tubes in three basket assembly,It can be controlled respectively.
Auto stop at timing point.
The bath liquid can achieve an even temperature.
A MPU is used to automatically control the temperature,time and frequency of nacelle back and forth.
Timing at will,The preset and real time data can be displayed alternately in time.
Automation:auto-test,auto-diagnose,auto-alarm.
Technical data:
Basket Assembly: 3
Nacelle back and forth frequency:30-32/min
Nacelle back and forth range :55+/-1 mm
Automatic timing range: 5-120 min
The range of temperature control: room temperature- 45 c
The accuracy of temperature control: +/- 0.5 c
Power 220V/50HZ/600W or 110V/60HZ/600W
Dimension: 540*300*420 mm
BJ-3 Disintegration tester