Air Leakage Tester for Positive Pressure Subjected Infusion Sets
1.The tester is designed and manufactured according to GB 8368-2005 (ISO 8536-4:2004 “infusion equipment for medical use-Part 4: infusion sets for single use, gravity feed, MOD”.) It applies an internal air pressure of 50kpa above atmospheric pressure for 15s to examine the infusion set for air leakage.
2.Range of positive pressure output: adjustable from 120kpa to 160kpa containing local Atmospheric pressure; with LED digital display; error: within±5%kpa.
3.Positive pressure reading displayed on the manometer: shall not fall 1kpa for 1 minute.
4.Duration: adjustable from 1s to 60s; with LED digital display; error: within ±1s.