Thermocouple Temperature Process Calibrator :
***Basic Functions:
♦ Measure / simulate /source output thermocouple: -200 ~1800°C (-328 °F ~ 3272 °F) ± (0.3°C + 10µV)
♦ Measure / simulate mV: -10 ~ 75mV ±(0.02%+2Lsd) ***Special Functions: ♦ A precise source and measurement tool for calibrating thermocouple instruments through a thermocouple mini jack. can be indicated in units of °C, °F , or mV.♦ Measure temperature from TC output♦ Simulate TC output♦ Operable with eight types of thermocouples♦ Calibrate linear TC transmitter with mV source function♦ Selectable °F or °C♦ Secondary injection
♦ Clear LCD max.display 99999 counts with backlight
***Environment Characteristics:
♦ Operating Temperature : -10ºC ~ 55ºC ♦ Storage Temperature : -40ºC ~ 60ºC
***General Characteristics:
♦ Power: 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)♦ Product color: Yellow & Black
♦ Product size : 190 * 89 * 42mm♦ Product Net Weight: 350g
♦ Standard Accessories: Test Leads, battery, User Manual, Aluminium Carry Case
♦ Standard Individual Packing: Aluminium Carry Case
♦ Standard Quantity Per Carton: 10pcs
♦ Standard Carton Meas: 58*46*22cm
♦ Standard Carton Gross Weight: 14.0kg
***Applications:
♦ Apply for testing basic electronic parameter
♦ Apply for calibrating the accuracy of the other instrument meter
♦ Apply for on-site process calibrating & controling in industrial field
***WARRANTY: one year
--------------------------------------------------
Technical Specifications:
1. Thermocouple Properties:
Thermocouple Type | Temperature Ranges | Display Resolution |
J | -200~1200 °C or -328~2192 °F | 0.1 °C or °F |
K | -200~1370 °C or -328~2498 °F | 0.1 °C or °F |
T | -200~400 °C or -328~752 °F | 0.1 °C or °F |
E | -200~950 °C or -328~1742 °F | 0.1 °C or °F |
R | -20~1750 °C or -4~3182 °F | 1 °C or °F |
S | -20~1750 °C or -4~3182 °F | 1 °C or °F |
B | 600~1800 °C or 1112~3272 °F | 1 °C or °F |
N | -250~1300 °C or -418~2372°F | 0.1 °C or °F |
2. Millivolt Measure and Source (Max input Vol:30v):
Range | Resolution | Accuracy |
-10mV~75mV | 0.01mV | ±(0.02%+2Lsd) |
3. Temperature Measure and Thermocouple Simulate(Max input vol:30v):
Thermocouple Type | Resolution | Accuracy |
J, K, T, E, N | 0.1° C or °F | ±(0.3 °C+10μV) |
R, S, B | 0.1 °C or °F | ±(0.3° C+10μV) |
--------------------------------------------------
---------------------------------------------