Introduction
The Tesla meter is a special instrument used for inspecting & checking flux density ,which is also one of the most universally devices in the field of magnetic measurement.
Model WT10C digital Tesla meter is controlled by the SCM, which is suitable for handle operation.It can be used to measure DC or AC magnetic field and flux dencity.The device can be carried on one’s person. It is characterized by its wide measuring range,simple operation and clear display. It is still added with the function of maintenance such as holding Measure Value/Peak Value, As mT or Gs unit of display can be change over, Measuring range of 200mT or 2000mT can be chose, and others that as resetting zero by key and so on. The power is one piece of battery 9V. It can be used continually about 20 hours.
Reality Working Domain:
1) Magnetic field distribution of material surface
2) Working in the magnetic field produced from some instrumentssuch as iron-removing device,magnet-selecting machine, magnet chuck, magnet spool and demagnetizing coil
3) Leakage Magnetic iron-surface
Main Technical Parameters
Rang: 0~200mT~2000mT
Accuracy: ±5%
Sensitivity: 0.01mT, 0.1mT
AC Magnet Field: 50-200Hz
Temperature of environment: 5°C-40°C
Relative humidity: 20%-80%(no condensation)
Power Supply: Battery 9V, constant DC power
Dimension: 150×70×30mm
Weight: N.W: 500g G.W:900g
Display: 4½ LCD
Shape of the device:
Sensor ConnectingLED ScreenZero/Reset ReadingPeak/Real Value Switch
DC/AC Switch
Range Changing
mT/Gs Unit Changing
Power On/Off
External power supply socketScreen of Display
Manipulation Procedure
1. Put the 9V battery to case which is in the back of the instrument; If the outer power is used, connect the terminal of the power to the socket which is in the side of the instrument and put another plug to the electric supply.
2. Insert the Hall probe into the socket on the panel (according to the arrow marking the sensor).
3. Press the key of the Power On/OFF in the panel and the digital display screen should display zero.
Select a suitable measurement range
4. Press the key RANGE switchover again and again , the measurement range would be transformed within 0-200 mT or 0-2000 mT.
5. Select DC/AC mode
Press the key DC/AC switchover again and again , the DC/AC mode would be selected.
6. Select display unit
Press the key(5) of conversion , the display unit would be transformed from mT to Gs.
7. Reset zero
Keep the sensor far away from magnetic field what it is not in the state of holding .If the display screen should not show “000”.you must press the key(7) to adjust it to zero.
Note: After change over the measurement range or transform the AC/DC mode , you all need to reset zero.
8. Measurement of Peak-value holding
Press the key(6) to change over normal measurement or Peak-value holding .
9. Loosen the cap of the Hall probe (see upper Fig).Place the effective position of Hall probe to touch on the surface of the resting material tightly (DC magnetic field) or at zone of measured magnetic field. The screen(8) will show the value of the magnetic field.
10. Reset Peak-value holding
When keeping the Peak-value holding , you need to measure new peak value and the forecast value will be lower forecast. Thus you could press the Reset button(7) to make it show a new peak value.
11. If the measurement is finished, the cap should be screwed up. It may be better to remove the battery from the case to prolong its time of use.
12. The Hall probe can be replacable .The difference from original probe to others is 2%.
Note: 0.1mT=1Gs 1T=10000Gs
Mistakes:
1,Don’t attempt to repair, or disassemble, and any reconstruct this instrument.
2, Use only recommended power accessories.
3, Don’t connect the terminals of the outer power supply reversely. the inner is plus, the outer is minus.
4, Don’t drop this instrument or the probe down ground. It’ll be damaged by strong vibration.
5, Do not allow hands or other objects with dirt to touch the probe.
6, Avoid using, placing or storing the instrument in places subject to strong sunlight or high temperature, as well as humidity, water, oil, rust air, vibration and others.
7, It would be used and stored up in temperature range
8, You should not move the instrument rapidly between high and low temperature to avoid condensation.
9, Never place the instrument in the places close to the machine generating strong magnetic field, such as magnet-charger, demagnetize coil, microwave oven, electric welding equipment electric motor and etc.
10,In conclusion, we suggest you calibrate the instrument in time.