APPLICATIONS
SP-3400 Gas Chromatograph, made according to international standards, its technical feature and index are in the world high level. The key components are imported from world famous suppliers. It is widely used for petrol-chemistry, environmental protection, epidemic prevention, pharmacology, scientific research, etc.
FEATURES
Ÿ Self-diagnostic function
1) Core tests
2) Automatic tests
3) Extended tests
4) Basic tests
Monitor the status of GC continuously. Once finding faults, it’ll display the information and show the wrong zone and the settling method.
Ÿ Self-protective function
1) Overrun temperature protection
2) Short circuit hint
3) TCD filament protection
4) FID flameout hint
5) PFD exposed-light protection
6) Keyboard locking with password
etc., ensuring normal running.
Ÿ Simple operation, powerful automation
1) All parameters can be entered through keyboard with prompt function
2) 4 sets of complete chromatography analysis methods can be stored and recalled automatically
3) An autosampler can be connected
4) Parameters can be modified instantly while the GC is running
5) Chromatography analysis method can be activated for 99 times repeatedly at the same time. It’s especially suitable for unattended operation.
Ÿ More choices of injectors
1) On- Column Injector for Packed Column
2) Flash Vaporization Injector for Packed Column
3) Automatic or manual gas inject valve
4) Headspace sampler
5) Thermal desorption system
6) Split/Splitless Capillary Injector
Three Injectors or two Split/Splitless Capillary Injectors can be fixed on GC.
Ÿ More choices of detectors
1) TCD
2) FID
3) ECD
4) FPD
5) TSD
Maximum of two TCDs or three different kinds of detectors can be installed.
Ÿ Reactor
1) Internal
2) external
Ÿ Time programming of detectors
Each of detectors has 5-ramp programmable time control. Output-signal, attenuation range, and polarity can be set automatically.
Ÿ Time programming of external events
Providing 4 external events with 20-ramp programmable time control. The optional GC relays may be used to automate valves, operate split/splitless capillary injectors, drive auxiliary devices, or switch signals between detector A and detector B in a run.
Ÿ Many kind of special-purpose GC can be provided according to user’s request.
SPECIFICATIONS
Ÿ Column oven
Operating temperature range ambient temperature to 420°C
Dual opening-door at the back of column oven
Cooling time 250°Cto 50°Cin 5 min
Programmable oven with 4-ramp temperature control
Temperature programming rate 0.1~50°C/min
Ÿ Injector system
Operating temperature range: ambient temperature to 420°C
Ÿ Thermal conductivity detector -TCD
Maximum temperature 420°C
Sensitivity ≥12000mv • mL/mg (for butane)
Linear range 105
Current range 50 to 400mA
Ÿ Filament protection
If He or H2 carrier gas flow in the detector cell is cut off for 4 minutes, filament power will be shut off automatically.
Ÿ Flame ionization detector-FID
Maximum temperature 420°C
Minimum detectivity ≤1.3×10-12g/s (n-C16)
Linear range 107
Flameout hint
monitoring the status of GC continuously after being turned on.
Ÿ Electron capture detector-ECD
Maximum temperature: 420°C
Minimum detectivity
≤0.1pg/mL (γ-666)
Linear range 104
Emission source: 11mC63Ni
Ÿ Flame photometric detector-FPD
Maximum temperature 420°C
Minimum detectivity
[P] ≤1×10-12g/s (tributyl phosphate)
[S] ≤1×10-10g/s (p-methylsulfurphosoharous)
Linear range
[P]105
[S]103
Two air-hydrogen flames For large volume injection or trace sample injection analyses
Single air-hydrogen flame sulfur sensitivity can be increased
Direct output for phosphorus or sulfur analyses
Square-root output for sulfur analyses
Exposed-light protection
When exposed-light current is more than 12μA, high voltage will be cut off automatically.
Ÿ Nitrogen-phosphorus detector-TSD
Maximum temperature: 420°C
Minimum detectivity
[N] ≤1×10-13g/s (azobenzene)
[P] ≤5×10-14g/s (malathion)
Linear range
[N]105
[P]104
SP-3400 Gas Chromatography