Technical specifications:
DEFIBRILLATOR:
(1) Defibrillator type: Manual, Synchronized, Asynchronized,
(2) Out put waveform: Monophasic technology /Biphasic technology (optional)
(3) Energy accuracy ∠+1%
(4) Energy selection 0,3,5,7,10,20,30,50,100,200,300,360 joule(nominal at a resistance of 50)
(5) Charging time Maximum 8 seconds to 360 Joules
(6) Standard Adult/Pediatric Paddles: Reusable extermal adult paddles(pediatric paddles integrated)
ECG:
ECG lead: Ⅰ, Ⅱ, Ⅲ, AVR, AVL, AVF, V1~V6, GND S-T segment analysis: -2.0~2.0mv Arrhythmia analysis: yes Protegtion: With stand 4000vAc/50Hz voltage in isolation and work against eletrosurgical interence and defibrillation
Respiration: Measurement method: Thoracic impedance
NIBP: Working mode: Manual/Automatic/stat(5 minutes) Method: Automatic oscillation Cufftype: standard Adult; optional Pediatric/Neonatal
Temperature: Measurement range 20~45°C
Resolution 0.1°C
SpO2:
Display: Spo2 value,pulse histongram,waveform,Pulse
Range: 0~99%for adult,pediatric and neonatal patients
Probe: standard adult finger clip
optional pediatric Y-type clip and neonatal wrap
Pulse range: 0-255bpm
Power requirements:
(a) AC Power supply 110V, 60 Hz /220V,50 Hz
(b) Vehicle Voltage DC 12V
Internal battery backup:
(1) Battery type: Rechargeable lithium battery.
(2) Charging time: Time-Minimum of 4 hours for full charging.
(3) Battery charging: During AC Power Supply operation of the device.
(4) Charging /Low Battery Indication: Red LED.
(5) Battery Capacity: 100% of the battery will provide 120 mins of monitoring; and 30 discharges at 360 joules.
Display: LED Dimensions: 7” diagonal {155mm*88mm}
Type: High resolution true-color graphic LED
Resolution: 480X234
Physical characteristics: Main unit: Portable
Standard parameter & configuration:
ECG/ RESP/SPO2/TEMP/PR/NIBP/DEFIBRILLATOR
Optional parameter: Recorder
Defibrillator PT-9000C