Functional Features
Advanced pressure steam way of sterilization internationally recognized.Convenient operating menus, customers are able to set up day and time etc.New generation structure of fully automatic electronic lock.Premium quality stainless steel stretched internal liner.High-luminance liquid crystal display.Built-in temperature sensing printer.Warm-up key setup, in order to keep the temperature in sterilization room around 70°C.(This function can reduce the cycle time by quickly reaching the sterilization temperature.)Build-in set by itself function with wide range of procedures and items available.Build-in cleaning and sewage equipment.Connect directly with external distilled water system or sewage pipe.Application Range
The autoclaves is mainly applied to operating room, dentistry, ophthalmology and biomedical research institutes, it can sterilize substances such as operation instruments, stomatological instruments, dressing, glassware, linjecting apparatus and others pressure steam resistant instruments.
Pre-set procedure | Tem(°C) | Pressure(bar) | Time of sterilization(Min) | Working total time (Min) | Loading | Maximum loading(kg) | |
1 | Prepare Vacuum three times 134°C Porous | 134 | 2.1 | 4 | 35-45 | Un-packed multi-loop equipment | 1.00 |
packed textile | 0.65 | ||||||
One-layer packing soild or vacancy substances | 3.00 | ||||||
Double-layer packing soild or vacancy substrnces | 1.50 | ||||||
2 | Prepare Vacuum three times121°C Porous | 121 | 1.0 | 20 | 56-60 | Un-packed multi-loop equipment | 1.00 |
Packed textile | 0.65 | ||||||
One-layer packing soild or vacancy substances | 3.00 | ||||||
Double-layer packing soild or vacancy substrnces | 1.50 | ||||||
3 | Prepare Vacuum one time134°C Hollow | 134 | 2.1 | 4 | 25-35 | Un-packed half-solid equipment | 6.00 |
4 | Prepare Vacuum one time121°C Hollow | 121 | 1.0 | 20 | 40-50 | Un-packed half-solid equipment | 6.00 |
5 | Prepare Vacuum one time134°C Solid | 134 | 2.1 | 4 | 20-30 | Un-packed solid equipment | 6.00 |
6 | Prepare Vacuum one time121°C Solid | 121 | 1.0 | 20 | 35-45 | Un-packed solid equipment | 6.00 |
7 | Prepare Vacuum three times134°C Prion | 134 | 2.1 | 18 | 45-60 | Un-packed multi-loop equipment | 1.0 |
Packed textile | 0.65 | ||||||
One-layer packing soild or vacancy substances | 3.00 | ||||||
Double-layer packing soild or vacancy substrnces | 1.50 | ||||||
8 | Prepare Vacuum three timesCustomer definition | 126°C(105°C~134°C) | - | 12min(1-40min) | Dryness time 15 minutes(3-25min) | - | |
9 | B-D Test | Applied to Performance Test | |||||
10 | Leak Test | Applied to Performance Test |
Dental Autoclave Class B 23 L